Ajay Prasad Hrishi

Ajay Prasad Hrishi

Profile ID RP-000-6695 Joined 2023-01-23 1 article
2
Total Views
1
Total Publications
2
Publication Views

Articles(1)

Authors: Karen Ruby Lionel, Ajay Prasad Hrishi

DOI: 10.18231/2394-4994.2018.0117  

Volume5

Issue4

Pages625-626

Abstract

High airway pressure’s post intubation is a critical event which requires the urgent intervention of the attending anesthesiologist. We report a case of high peak pressure and plateau pressure after intubation which was actually due to the kinked scavenging tubing of anesthesia workstation. Many a times this may happen, but is missed by even the experienced anesthesiologist who attribute the high airway pressures to change in the lung compliance of the patient or faulty equipment and will immediately respond with therapeutic intervention or replace the workstation/ventilator. A systematic evaluation of the lungs as well as the function of the ventilator along with the scavenging tubing’s, a commonly missed entity is mandatory in each case. Keywords: Airway pressure, Anesthesia, Scavenging system.